为贯彻《环境保护法》和《大气污染防治法》,保护环境,保障人体健康,规范固定污染源废气中挥发性有机物的测定方法,制定本标准。本标准规定了固定污染源废气中挥发性有机物的固相吸附-热脱附/气相色谱-质谱法。本标准为发布。
组合2吸附管,内装CarbopackB、Carboxen1000,长度分别为30、10mm。
5.7 吸附管:不锈钢材质,内径 5 mm。
组合 1 吸附管,内装 Tenax GR、Carbopack B,长度分别为 30、25 mm。
组合 2 吸附管,内装 Carbopack B、Carboxen 1000,长度分别为 30、10 mm。
C.2.2 组合 2 吸附采样管
适用于 C3~C12 范围内的化合物。在湿度低于 65%、温度低于 30℃时(含湿量
<1.8%),可采集气体样品 2 L;在湿度高于 65%、温度高于 30℃时(含湿量>1.8%),采
样体积要降到 0.5 L;对于 C4 以上的化合物,采样体积可增加到 5 L。老化温度 350 ℃,
老化流量 100 ml/min
Combination 2 adsorption tubes
Combination 2 adsorption tubes Details:
This standard is formulated in order to implement the Enviro
nmental Protection Law and the Air Pollution Prevention and Co
ntrol Law, protect the environment, safeguard human health, and standardize the determination methods of volatile organic compounds in waste gas from statio
nary pollution sources. This standard specifies the solid-phase adsorption-thermal desorption/gas chromatography-mass spectrometry method for volatile organic compounds in waste gas from statio
nary pollution sources. This standard is published for the first time.
HJ 734-2014 Waste gas from statio
nary pollution sources-Determination of volatile organic compounds-Solid phase adsorption-thermal desorption-Gas chromatography-mass spectrometry
Combination of 2 adsorption tubes, with CarbopackB and Carboxen1000 inside, the lengths are 30 and 10mm respectively.
5.7 Adsorption tube: stainless steel, inner diameter 5 mm.
Combination 1 adsorption tube with Tenax GR and Carbopack B inside, lengths of 30 and 25 mm, respectively.
Combination of 2 adsorption tubes with Carbopack B and Carboxen 1000 inside, lengths of 30 and 10 mm respectively.
C.2.2 Combination 2 adsorption sampling tube
Suitable for compounds in the range of C3 to C12. When the humidity is lower than 65% and the temperature is lower than 30°C (moisture content
<1.8%), 2 L of gas samples can be collected; when the humidity is higher than 65% and the temperature is higher than 30℃ (moisture content>1.8%),
The sample volume is reduced to 0.5 L; for compounds above C4, the sample volume can be increased to 5 L. Aging temperature 350 ℃,
Aging flow 100 ml/min