产品名称: |
HH-BLQ 药典二氧化碳玻璃球色谱柱研制 |
产品型号: |
HH-BLQ |
品牌: |
1941 |
产品数量: |
|
产品单价: |
面议 |
日期: |
2021-12-13 |
HH-BLQ 药典二氧化碳玻璃球色谱柱研制的详细资料
药典二氧化碳玻璃球色谱柱研制
药典二氧化碳玻璃球色谱柱研制 详细信息:
名称:药典二氧化碳(CO2)净化装置
型号:HHO-III Purification device
应用:2010版及2015版,2020版中国药典二氧化碳中碳氢化合物测定 气相色谱法
碳氢化合物照气相色谱法(通则0521)测定供试品气体取本品,即得对照品气体取甲烷含量为0.0020%的气体(以氮气为稀释剂)。色谱条件用玻璃球为填料的色谱柱(4mm×0.8m,80目);柱温为110℃;进样口温度为110℃;检测器为火焰离子化检测器,温度为120℃测定法精密量取供试品气体与对照品气体,分别注入
气相色谱仪,在净化温度为360℃时测得的峰面积为相应空白值;精密量取供试品气体与对照品气体,分别注入气相色谱仪,测定峰面积,减去相应空白值后的峰面积为校正峰面积限度按外标法以校正峰面积计算,含碳氢化合物(以甲烷计)不得过0.0020%。
浩瀚色谱(山东)应用技术开发有限公司,,依据仪器的规格型号,研发出适合安捷伦7890B气相色谱仪的1m*1/8的英寸玻璃球色谱柱,安装简单快速,结果准确。
药典二氧化碳玻璃球色谱柱研制 测试谱图:




.jpg)






Development of Chromatographic Column for Carbon Dioxide Glass Ball in Pharmacopoeia
Development of Pharmacopoeia Carbon Dioxide Glass Sphere Chromatography Column Details:
Name: Pharmacopoeia carbon dioxide (CO2) purification device
Model: HHO-III Purification device
Application: 2010 and 2015 editions, 2020 edition Chinese Pharmacopoeia Determination of hydrocarbons in carbon dioxide gas chromatography
Hydrocarbons are determined according to gas chromatography (General Principle 0521). Take this product as the test sample gas, that is, the reference sample gas should be a gas with a methane co
ntent of 0.0020% (with nitrogen as a diluent). Chromatographic co
nditions use a chromatographic column filled with glass balls (4mm×0.8m, 80 mesh); the column temperature is 110℃; the temperature of the injection port is 110℃; the detector is a flame io
nization detector, and the temperature is 120℃. The method is precise. Measure the test gas and reference gas and inject them into the gas chromatograph respectively. The peak area measured when the purification temperature is 360°C is the correspo
nding blank value; accurately measure the test gas and reference gas and inject them into the gas chromatograph. The peak area is determined by the instrument, and the peak area after subtracting the correspo
nding blank value is the calibrated peak area limit. The peak area is calculated by the external standard method, and the hydrocarbon co
ntent (calculated as methane) shall not exceed 0.0020%.
Haohan Chromatography (Shandong) Application Technology Development Co., Ltd. keeps pace with the times and develops a 1m*1/8 inch glass bulb column suitable for Agilent 7890B gas chromatograph according to the specifications of the instrument. The installation is simple and fast, and the results are accurate and reliable. .
Development of Pharmacopoeia Carbon Dioxide Glass Sphere Chromatographic Column Test Spectrum: