产品中心
产品展示

  您当前的位置:首页 » 产品展示 » 专用填充柱 » HH-C 活性炭测环境空气和废气中三甲苯

产品名称: HH-C 活性炭测环境空气和废气中三甲苯
产品型号: HH-C
品牌: 1941
产品数量:
产品单价: 面议
日期: 2023-12-14

HH-C 活性炭测环境空气和废气中三甲苯的详细资料

活性炭测环境空气和废气中三甲苯
活性炭测环境空气和废气中三甲苯 详细信息:

活性炭采样管:
规格:6*80mm100/50mg,6*120mm100/50mg,6*150mm100/50mg,6*180mm100/50mg,1/4*80MM
溶剂解吸型:CS2/二硫化碳
应用: 环境空气和废气 三甲苯的测定 活性炭吸附/二硫化碳解析-气相色谱法

滕州市浩瀚色谱仪器技术服务有限公司,建立了毛细管柱气相色谱法测定环境空气和废气中三甲苯,三甲苯经活性炭吸附/二硫化碳解吸、HH-WAX毛细管柱分离、氢火焰离子化检测器检测.研究结果表明,本方法测定1,3,5-三甲苯、1,2,4-三甲苯和1,2,3-三甲苯的标准曲线的相关系数均为0.9999;当采样体积为30 L、进样量为2.0μL时,方法度为0.24%—0.39%、加标回收率为96.7%—108.3%、1,3,5-三甲苯方法检出限为8.4×10^-4mg·m^-3、1,2,4-三甲苯方法检出限为6.9×10^-4mg·m^-3、1,2,3-三甲苯方法检出限为7.5×10^-4mg·m^-3.测试表明方法经优化性能后能有效去除干扰物的影响,采集后的样品可在4℃低温环境中保存7 d.

活性炭测环境空气和废气中三甲苯 测试谱图:


Activated carbon for measuring xylene in ambient air and exhaust gas
Activated carbon measurement of xylene in ambient air and exhaust gas Details:

Activated carbon sampling tube:
Specifications: 6 * 80mm100 / 50mg, 6 * 120mm100 / 50mg, 6 * 150mm100 / 50mg, 6 * 180mm100 / 50mg, 1/4 * 80MM
Solvent desorption type: CS2 / carbon disulfide
Application: Ambient air and exhaust gas Determination of xylene Activated carbon adsorption / carbon disulfide analysis-gas chromatography

Tengzhou Haohan Chromatographic Instrument Technology Service Co., Ltd. established capillary column gas chromatography for the determination of xylene in ambient air and exhaust gas. Xylene was detected by activated carbon adsorption / carbon disulfide desorption, HH-WAX capillary column separation, and hydrogen flame ionization detector. The results show that the correlation coefficients of the standard curves for the determination of 1,3,5-trimethylbenzene, 1,2,4-trimethylbenzene, and 1,2,3-trimethylbenzene are 0.9999; When the injection volume is 2.0 μL, the method precision is 0.24% -0.39%, the standard recovery is 96.7% -108.3%, and the detection limit of the 1,3,5-trimethylbenzene method is 8.4 × 10 ^ -4mg · m ^ -3,1,2,4-xylene method detection limit is 6.9 × 10 ^ -4mg · m ^ -3,1,2,3-xylene method detection limit is 7.5 × 10 ^ -4mg · m ^ -3. Tests show that the method can effectively remove the influence of interferences after optimizing the performance. The collected samples can be stored for 7 d in a low temperature environment of 4 ° C.

Measurement of trimethylbenzene in ambient air and exhaust gas with activated carbon